Promass P 100 offers the enhanced performance of a bent tube meter and the drainability of a straight tube meter without any compromise. Combined with the smallest transmitter housing available today it delivers full performance on the smallest footprint. Designed for applications in sterile environments where space is a premium, Promass P 100 will be the preferred choice for system integrators, skid builders and equipment manufacturers.
Measuring principle operates independently of physical fluid properties such as viscosity or density. Dedicated for applications under sterile conditions in the life sciences industry. Specialist for life sciences with an ultra-compact transmitter. Dedicated to applications under sterile conditions in the life sciences industry.
DN 8 to 50 (⅜ to 2″)
Measuring tube: 1.4435 (316L)
Connection: 1.4435 (316L); 1.4404 (316/316L)
Mass flow, density, temperature, volume flow, corrected volume flow, reference density, concentration
Mass flow (liquid): ±0.1 %
Volume flow (liquid): ±0.1 %
Density (liquid): ±0.0005 g/cm³
0 to 70 000 kg/h (0 to 2570 lb/min)
PN 63, Class 300, 40K
Standard: –50 to +150 °C (–58 to +302 °F)
Option: –50 to +205 °C (–58 to +401 °F)
Standard: –40 to +60 °C (–40 to +140 °F)
Option: –50 to +60 °C (–58 to +140 °F)
1.4301 (304), corrosion resistant
Compact: AlSi10Mg, coated
Compact/ultra‐compact: 1.4301 (304)
Standard: IP66/67, type 4X enclosure
Option: IP69
4‐line backlit display available (no local operation)
Configuration via web browser and operating tools possible
4‐20 mA HART (active)
Pulse/frequency/switch output (passive)
None
HART, Modbus RS485, EtherNet/IP, PROFIBUS DP, PROFINET
DC 20 to 30 V
ATEX, IECEx, cCSAus, INMETRO, NEPSI, EAC
CE, C-Tick, EAC marking
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
Heartbeat Verification: Heartbeat Technology complies with requirements for traceable verification according to ISO 9001:2008, chapter 7.6. a (TUV attestation)
PED, CRN
3.1 material
3-A, EHEDG, ASME BPE, ISPE, cGMP
Highest process quality – fully compliant to industry requirements. Fewer process measuring points – multivariable measurement (flow, density, temperature). Space‐saving installation – no in-/outlet run needs. ASME BPE, 3‐A and EHEDG conform & low delta ferrite. Electropolished measuring tube in 1.4435 (316L). Fast recovery from CIP/SIP.
Space‐saving transmitter – full functionality on the smallest footprint. Time‐saving local operation without additional software and hardware – integrated web server. Integrated verification – Heartbeat Technology. Robust, ultra-compact transmitter housing. Highest degree of protection: IP69. Local display available.