Proline Prowirl R 200 vortex flowmeter

Prowirl R was designed for low flows and is thus the particularly dependable solution for energy management. Additionally its calibration option PremiumCal guarantees excellent measuring accuracy for maximum plant availability. With genuine loop-powered technology, Prowirl R 200 enables cost-effective and seamless integration into existing infrastructures. It offers highest operational safety in hazardous areas. Heartbeat Technology enables process safety at all times.

Flowmeter with best-in-class accuracy despite pipe reduction. Easy energy management – integrated temperature and pressure measurement for steam and gases. Dedicated to applications with very low flow or reduced flow. Preferred measuring principle for wet/saturated/superheated steam, gases & liquids (also cryogenic). Dedicated to applications with very low or reduced flow.

Device properties

  • Integrated diameter reduction by 1 or 2 line sizes
  • Nominal diameter (mating pipe) up to DN 250 (10″)
  • Flexible positioning of pressure cell
  • Display module with data transfer function
  • Robust dual-compartment housing
  • Plant safety: worldwide approvals (SIL, Haz. area)

Benefits

  • Easy energy management integrated temperature and pressure measurement for steam and gases
  • Cost and time savingsno pipework modifications needed for line size reduction
  • Same accuracy down to Re 10 000 – most linear Vortex meter body
  • Long-term stabilityrobust drift-free capacitive sensor
  • Convenient device wiringseparate connection compartment, various Ethernet options
  • Safe operationno need to open the device
  • Integrated verificationHeartbeat Technology

Nominal diameter range

DN 25 to 250 (1 to 10″)

Wetted materials

Measuring tube: 1.4408 (CF3M)

DSC sensor: 1.4404/F316/F316L

Process connection: 1.4404/F316/F316L

Measured variables

Volume flow, mass flow, corrected volume flow, energy flow, heat flow difference, temperature

Max. measurement error

Volume flow (liquid): ±0.75 %

Volume flow (steam, gas): ±1.00 %

Mass flow (saturated steam): ±1.7% (temperature compensated); ±1.5% (temperature/pressure compensated)

Mass flow (superheated steam, gas): ±1.5 (temperature/pressure compensated); ±1.7% (temperature compensated + external pressure compensation)

Mass flow (liquid): ±0.85%

Measuring range

Liquid: 0.2 to 540 m³/h (0.15 to 320 ft³/min)

depending on medium: water with 1 bar a, 20 °C (14.5 psi a, 68° F)

Steam, gas: 1.5 to 7260 m³/h (0.9 to 4270 ft³/min)

depending on medium: steam with 180 °C, 10 bar a (356 °F, 145 psi a); air with 25 °C, 4.4 bar a (77 °F, 63.8 psi a)

Max. process pressure

PN 40, Class 300, 20K

Medium temperature range

Standard: –40 to +260 °C (–40 to +500 °F)

High/low temperature (option): –200 to +400 °C (–328 to +752 °F)

High/low temperature (on request): –200 to +450 °C (–328 to +842 °F)

Ambient temperature range

Compact version (standard): –40 to +80 °C (–40 to +176 °F)

Compact version (option): –50 to +80 °C (–58 to +176 °F)

Remote version (standard): –40 to +85 °C (–40 to +185 °F)

Remote version (option): –50 to +85 °C (–58 to +185 °F)

Sensor housing material

Sensor connection housing: AlSi10Mg, coated; 1.4408 (CF3M)

Transmitter housing material

AlSi10Mg, coated; 1.4404 (316L)

Degree of protection

Compact version: IP66/67, type 4X enclosure

Sensor remote version: IP66/67, type 4X enclosure

Transmitter remote version: IP66/67, type 4X enclosure

Display/Operation

4‐line backlit display with touch control (operation from outside)

Configuration via local display and operating tools possible

Remote display available

Outputs

4‐20 mA HART (passive)

4‐20 mA (passive)

Pulse/frequency/switch output (passive)

Inputs

4‐20 mA (passive)

Digital communication

HART, PROFIBUS PA, FOUNDATION Fieldbus

Power supply

DC 12 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)

DC 12 to 30 V (4‐20 mA HART, 4‐20 mA)

DC 12 to 35 V (4‐20 mA HART, pulse/frequency/switch output, 4‐20 mA input)

DC 9 to 32 V (PROFIBUS PA, pulse/frequency/switch output)

Hazardous area approvals

ATEX, IECEx, cCSAus, JPN, EAC, UK Ex

Product safety

CE, C-tick, EAC

Functional safety

Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

Metrological approvals and certificates

Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)

Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)

Marine approvals and certificates

ABS, LR, BV, DNV

Pressure approvals and certificates

PED, CRN, AD 2000

Material certificates

3.1 material

NACE MR0175/MR0103, PMI (on request); welding test acc. to ISO 15614‐1, similar to ASME IX (on request)

Sensor features

Cost and time savings – no pipework modifications needed for line size reduction. Same accuracy down to Re 10 000 – most linear Vortex meter body. Long-term stability – robust drift-free capacitive sensor. Integrated diameter reduction by 1 or 2 line sizes. Nominal diameter (mating pipe) up to DN 250 (10″).

Transmitter features

Cost and time savings – no pipework modifications needed for line size reduction. Same accuracy down to Re 10 000 – most linear Vortex meter body. Long-term stability – robust drift-free capacitive sensor.

Integrated diameter reduction by 1 or 2 line sizes. Nominal diameter (mating pipe) up to DN 250 (10″). Flexible positioning of pressure cell.