iTEMP TMT85 temperature transmitter

Unsurpassed reliability, accuracy and long-term stability in critical processes over all industries. The configurable transmitter not only transfers digital converted signals from RTD and TC sensors, it also transfers resistance and voltage signals using FOUNDATION Fieldbus™ communication. High measurement point availability by means of sensor monitoring functions and device hardware fault recognition. Optimization of the measurement accuracy by sensor-transmitter matching

FOUNDATION Fieldbus™ temperature transmitter as head or field device with two universal sensor inputs suitable for use in hazardous areas. 2 universal input channels and FF™ protocol for the conversion of input signals into digital output signals. The device stands out due to signal reliability, long-term stability, high precision and advanced diagnostics (important in critical processes). For the highest level of safety, availability and risk reduction. Installation in flat-face terminal head as per DIN EN 50446. Optional installation in field housings even for use in Ex d areas – with mounting brackets for wall or pipe mounting.

Benefits

  • Easy and standardized communication via FOUNDATION Fieldbus™ H1
  • Straightforward design of measuring points in Ex-areas through FISCO/FNICO conformity in accordance with IEC 600079-27
  • Safe operation in hazardous areas thanks to international approvals such as FM IS, NI; CSA IS, NI; ATEX Ex ia, Ex nA for intrinsically safe installation in zone 1 and zone 2
  • High accuracy through sensor-transmitter matching
  • Reliable operation with sensor monitoring and device hardware fault recognition
  • Several mounting versions and sensor connection combinations
  • Rapid no-tools wiring due to optional spring terminal technology

Measuring principle

Head transmitter

Input

2 x RTD, TC, Ohm, mV

Output

FOUNDATION Fieldbus H1

Auxiliary power supply

9…32 V DC (FOUNDATION Fieldbus)

9…17,5 V DC (FISCO/FNICO)

Communication

FOUNDATION Fieldbus H1

Installation

Terminal head form B

Accuracy

(Pt100) <= 0,1 K

(Pt100) <= 0,18 °F

Galvanic isolation

yes

Certification

ATEX Ex ic, Ex ia, Ex nA, Ex tc, Ex d

ATEX IECEx Ex ia

IECEx Ex ia, Ex tb, Ex d

CSA C/US General Purpose

CSA IS, NI, XP, DIP

FM IS, NI, XP, DIP

FM+CSA IS, NI

FM+CSA IS, NI + ATEX IECEx Ex ia

INMETRO Ex ia, Ex d, Ex tb

NEPSI Ex ia, Ex ec, Ex d, Ex tb

EAC Ex ia, Ex d

UKCA Ex tb, Ex tc, Ex ia, Ex nA, Ex db

Features

  • Dual-channel input – allows two independent temperature measurements.
  • Universal input types:
  • RTD sensors (Pt100, Pt200, Pt500, Pt1000, etc.) according to IEC 60751
  • Thermocouples (Types B, E, J, K, N, R, S, T, etc.)
  • Resistance (Ω) and voltage (mV) signals
  • Output signal: FOUNDATION Fieldbus H1 digital protocol
  • Accuracy: up to ± 0.1 K for Pt100
  • Measuring range: typically –200 °C to +850 °C (depending on sensor type)
  • Installation: fits DIN EN 50446 Form B sensor heads or field housings
  • Diagnostics and reliability:
  • Sensor monitoring for open/short circuits or cable corrosion
  • Channel redundancy: automatic switch to second channel on failure
  • Sensor drift detection and alarm functions
  • Ex protection: approved for hazardous (Ex d, Ex ia, etc.) zones