iTEMP TMT72 temperature transmitter

Unsurpassed reliability, accuracy and long-term stability in critical processes across all industries. The configurable transmitter transfers converted signals from RTD and thermocouple (TC) temperature probes, as well as resistance and voltage signals using HART® communication. Diagnostics information according to NAMUR NE 107. Seamless integration into all major control and asset management systems. Integrated Bluetooth® interface for remote operation via Endress+Hauser SmartBlue App.

HART® temperature transmitter as head, field or DIN rail device with one universal sensor input suitable for use in hazardous areas. Bluetooth®-enabled temperature transmitter with HART® communication for the conversion of various input signals into a scalable, analog 4 to 20 mA output signal. Improved process efficiency and increased plant uptime. Universal input for RTD, TC, resistance transmitters (Ω), voltage transmitters (mV). Installation in terminal head form B (flat face) as per DIN EN 50446. Installation in cabinets with DIN rail housing. Optional installation in field housing even for use in Ex d applications.

Benefits

  • Safe operation in hazardous areas thanks to international approvals
  • Reliable operation thanks to sensor and device monitoring
  • Integrated Bluetooth® interface for the wireless display of measured values and configuration via E+H SmartBlue (app), optional
  • Diagnostics information according to NAMUR NE 107
  • Available with plug-on display TID10

 

Measuring principle

Head transmitter

Input

1 x RTD, TC, Ohm, mV

Output

1 x analog 4…20 mA / HART

Auxiliary power supply

10…36 V DC (Head transmitter)

11…36 V DC (DIN rail device)

10/11…30 V DC (Ex ia)

Communication

HART-protocol

Bluetooth® wireless technology

Installation

Terminal head form B / DIN rail / field mount housing

Accuracy

(Pt100, -50…200 °C) <= 0,1 K

(Pt100, -58…392 °F) <= 0,18 °F

Galvanic isolation

yes

Certification

ATEX Ex ia, Ex ic, Ex tc, Ex nA, Ex db, Ex tb

ATEX IECEx Ex ia, Ex tb

IECEx Ex d, Ex tb

CSA C/US General Purpose

CSA C/US IS, Ex ec, XP, DIP

INMETRO Ex ia, Ex d, Ex tb

NEPSI Ex ia, Ex d, Ex tb

EAC Ex ia, Ex d

UKCA Ex ia, Ex tc, Ex nA, Ex db, Ex tb

Features

  • Universal input types
  • RTD: Pt100, Pt200, Pt500, Pt1000, etc.
  • Thermocouples: types J, K, T, E, N, R, S, B, C, D, etc.
  • Resistive sensors and voltage input (, mV) over appropriate ranges.
  • Typical accuracy: For Pt100 in the range –50…+200 °C the accuracy is 0.1 K.
  • Communication and configuration: HART® protocol, optional Bluetooth® interface for wireless display/configuration via the SmartBlue® app.
  • Wiring & installation: Push-in terminals optional for tool-free wiring.
  • Diagnostics: Sensor and device monitoring, diagnostics per NAMUR NE 107.
  • Power supply: 10-36 V DC (typical for head / DIN-rail versions) depending on variant.